+86-27-65526006

+86-27-65526007

Home> Products> ZX-3125A Insulation Resistance Tester

ZX-3125A Insulation Resistance Tester

Insulation testing is very important to ensure the quality of products and the safety of people and equipment in operation.

Before adding power to the system, the insulation test is used to verify that it is sound and can improve the performance of the system. Insulation test can find manufacturing process problems and equipment defects, which can not be found before the equipment fails. In the European Union, the test is mandatory, even for the smallest civilian systems.

Features
Technical indicators

1. Measuring range and accuracy(Temperature, humidity;23±5℃,45-75%RH)

Rated voltage

250V

500V

1000V

2500V

5000V

Measure Range

0.1~100.0MΩ

0.1~99.9MΩ

80~1000MΩ

0.1~99.9MΩ

80~999MΩ

0.80~2.00GΩ

0.1~99.9MΩ

80~999MΩ

0.80~9.99GΩ

8.0~100.0GΩ

0.1~99.9MΩ

80~999MΩ

0.80~9.99GΩ

8.0~99.9GΩ

80~20000GΩ

Display range

0.1~100.0MΩ

0.1~1000MΩ

0.1M~2.00GΩ

0.1M~100.0GΩ

0.1M~2000 GΩ

Open circuit Voltage

DC250V

+10%,-10%

DC 500V

+ 20%,-10%

DC 1000V

+ 20%,-0%

DC 2500V

+20%,-0%

DC5000V

+20%,-0%

Rated Current



1MΩ load

1mA~1.2mA

2.5MΩ load

1mA~1.2mA

5MΩ load

1mA~1.2mA

Short-circuit Current

5mA±0.5mA

Accuracy

±5%±3dgt

±5%±3dgt±20%

Voltage monitor for insulation resistance range

30 – 600V (resolution 10 V): ±10%rdg±20V

This monitor is used to check whether electric charges stored no the equipment under test are discharged or not. The voltage value measured and displayed on the LCD is a reference value. Please note that the indicated value, when an external AC voltage is applied to the instrument, is not a correct value.


DC Voltage

AC Voltage

Measuring range

±30 -- ±600V

30 -- 600V(50/60Hz)

Resolution

1V

Accuracy

±2%rdg±3dgt

Display

Liquid crystal display insulation resistance range:(Max.1200 counts) VAC/DC range: (Max.630counts) Bar graph: Max.36points DAR/PI value:Max.9.99 Time:Max.99.59

Low battery warning

Battery symbol (in 4 levels)

Over range indication

“OL” mark appears on insulation resistance range.

Auto-ranging

Range shifts to upper range:1000 counts

Range shifts to lower range:80 counts

(Only on the insulation resistance range)

Auto-power-off

The instrument will be powered off if there is no switch or button operation for 10 min.(This function does not work during a measurement.)

Altitude

2000m or less

Temperature & humidity range(guaranteed accuracy)

23℃±5℃/Relative humidity 85% or less(no condensation)

Operating temperature & humidity range

0℃ to 40 ℃/Relative humidity 85% or less (no condensation)

Storage temperature&humidity range

-20℃ to ± 60%/ Relative humidity75% or less(no condensation)

Overload protection: Insulation resistance range

AC 1200V/10sec.

Withstand voltage

Voltage range:AC720V/10sec.

AC5160V(50/60Hz)5sec.(Between electrical circuit and enclosure)

Insulation resistance

AC5160V(50/60Hz)5sec.(Between electrical circuit and enclosure)

Dimension

177(L)×226(W)×100(D)mm

Weight

1.9kg(battery included)

Power source

12.6V/6800mAh

2. Current consumption (representative values at 12V of supply voltage)

Range

250V

500V

1000V

2500V

5000V

VAC/DC

Output at Short-circuit

600mA

110mA

When rated current is outputted

650mA/0.25MΩ

800mA/0.5MΩ

1000mA/1MΩ

1350mA/2.5 MΩ

1350mA/5 MΩ

Output at open circuit

40mA

40mA

50mA

80mA

150mA

Stand-by

25mA

Backlight on

Increased by 40mA

3. Measurement time: for 25hours *under a load of 100MΩ

on the Insulation resistance 5000V range.

Service commitment
Baidu
map